
2
11-- IINNTTRROODDUUCCTTIIOONN
The frequency and nature of testing or calibration may be determined by local
regulation or standards.
EN378 and the FGAS Regulation require an annual check in accordance with
the manufacturer’s recommendation.
Murco recommends annual checks by bump test, and gas calibration
on site at two yearly intervals in the case of semiconductor SC sensors
and infrared IR sensors with sensor replacement every five years or as
required. In the case of electrochemical EC sensors we recommend
annual bump test and sensor replacement at two yearly intervals and
gas calibration. This should eliminate end of life concerns, and
constantly renew the detection system.
If the MGS is exposed to a large leak it should be tested to ensure correct
functionality by electrically resetting the zero setting and carrying out a
bump test, see procedures below.
There are two concepts that need to be differentiated: bump test and
calibration
Bump Test
: This consists of exposing the sensor to a gas and observing its
response to the gas. The objective is to establish if the sensor is reacting to
the gas and all the sensor outputs are working correctly. There are two types
of bump test.
Quantified
: where a known concentration of gas is used, or
Non-Quantified: where a gas of unknown concentration is used.
Calibration
: This consists of exposing the sensor to a calibration gas, setting the
“zero” or “Standby voltage”, the span / range, and checking/adjusting all the
outputs, to ensure that they are activated at the specified gas concentration.
Procedures for bump test and calibration vary depending on the sensor
technology used and the gas in question. The MGS is available in three sensor
versions: Semiconductor (
SSCC
), and Electrochemical (
EECC)) aanndd iinnffrraarreedd ((IIRR))
.
Before you carry out the test or calibration:
1- Advise occupants, plant operators, and supervisors.
2- Check if the MGS is connected to external systems such
as sprinkler systems, plant shut down, external sirens
and beacons, ventilation, etc. and disconnect as
instructed by the customer.
3- Deactivate alarm delays if selected at JP5, JP6 as per
instructions in Diagram 1.
4- For Bump Test or Calibration the MGS should be
powered up overnight. If the unit has been installed and
running for about 24 hrs, and you need to power it off
for a short time to set the delay at 0 min, then the
normalisation period is about 5 min (this is indicated by
the green LED flashing) and then you can begin the
testing or calibration. If sensors have been in long-term
storage or the detectors have been turned off for a long
time, normalisation would be much slower. However
within 1-2 hours the sensor should have dropped below
the alarm setting and be operational. You can monitor
normalisation progress exactly by monitoring the
sensor output, on CON 2 between pins OV & VS.
See Section 5.
22-- BBUUMMPP TTEESSTT ((eevveerryy yyeeaarr))
Ideally bump tests are conducted on site
in a clean air atmosphere.
Prior to carrying out a bump test, check
and adjust the zero setting as described
in the Calibration section page 3 -
Sensor PCB 1.1 and 2.2.
2.1 Semiconductor and IR sensors for
hydrocarbons: We offer cylinders of
gas at known concentrations for
quantified tests. This consists of
exposing the sensor to the gas and
checking that alarm lights and relays
are activated. If this is not available,
for a non-quantified test you can use
a gas cigarette lighter. By cracking
open the valve without igniting the
gas, you release the gas onto the
sensor and force it into alarm. Check
that alarm light, sounder and relay
are activated.
2.2 Electrochemical sensors: We offer Ampoules of ammonia (NH3) at 100ppm
and 1.000 ppm. Cylinders of calibration gas are also available. These are a
quantified test. For details of other Ampoules, please contact us.
2.3 Infrared sensor for CO2: We offer Ampoules of CO2 at 5000 ppm.
Cylinders of calibration gas are also available. These are a quantified test. If
these are not available, then you can breathe on the sensor. Human breath
has enough CO2 to trigger the alarm. This is a non-quantified test.
2.4 Bump test using gas ampoules:
1- Make sure that both the ampoules and the
calibration beaker are clean and dry.
2- Unscrew the beaker hold screw and place
the ampoule so that is sits in the
base of the beaker. As per
illustration.
3- Tighten on the screw
ampoule without breaking it.
4- Remove the enclosure lid of
the gas detector (not in Ex area).
5- Connect voltmeter to monitor sensor response, monitor 0-
10v (Jumper JP1 and JP3 off) response on CON 2
between pins OV & V.
6- Place the beaker over the sensor head using the multi
sensor adaptor to fit the sensor, or, if an Exd, IP66 or
Remote sensor head version, screw the beaker on the
remote sensor head M42 thread, or M35 thread adaptor. It
should be as tight fitting as possible to allow maximum
exposure to the gas.
Flexible non-
absorbant
tubing
Flow
Regulator
Typical Flow
0.3L/Min
Outlet 5/8’’
18 UNF
(C10)
Cylinder
360mm high x 88mm diameter
440mm high with flow regulator
110L capacity of calibration gas
Weight 1.4 kg
V
ented
Calibration
Hood
Calibration Kit
6
5
4
3
2
1
A
D
NC
COM
NO
I
V
0V
iv
0V
Relay
+ Sounder
P2
P1
P3
ZeroSpan
SW2 SW1
LD2 LD1
CN1 CN2 CN3
JP6
JP1
Relay & Sounder Delay – JP5 & JP6 off : no delay
JP5 on only : 1 Minute
JP6 on only : 5 Minutes
JP5 & JP6 on : 10 Minutes
JP3 – on : 4-20mA or 2-10Volt output
off : 0-10Volt or 0-5 Volt output
JP2 – on : Sounder enabled
JP1 – on : 0-5Volt output
off : 0-10Volt output
DDiiaaggrraamm 11
Comentarios a estos manuales